Haptic-Optic Fractures in Yamane Flanged Intrascleral Fixated Lenses: A Case Series
Published date: Sep 02 2025
Journal Title: Journal of Ophthalmic and Vision Research
Issue title: Volume 20 - 2025
Pages: 1 - 6
Authors:
Abstract:
Purpose: To report five cases of structural weakness at the haptic-optic junction of the CT Lucia 602 intraocular lens (IOL) (Carl Zeiss Meditec Inc., Dublin, CA), including two optic breakages and three optic microfractures, while using the Yamane flanged intrascleral haptic fixation surgery (FIHFS).
Methods: A retrospective chart review of five cases that underwent Yamane FIHFS by the same surgeon between 2020 and 2022 was performed. Patients’ demographics, operative technique, postoperative outcomes, and adverse events were recorded over a 12-month followup period.
Results: One case of optic breakage, noted on postoperative week 1 (POW1), had the optic with the trailing haptic suspended in the vitreous while both haptic flanges remained secured to the sclera, requiring IOL exchange. The second case of optic breakage was noted intraoperatively and required explantation of the IOL with subsequent placement of an iris-fixated IOL. Three cases of optic microfractures were observed with no long-term complications. All five fractures occurred at the junction of the leading haptic.
Conclusion: Optic breakage and microfractures at the haptic-optic junction are unusual complications of intrascleral fixated IOL using the Yamane technique. To the best of our knowledge, this has not been previously reported with the CT Lucia 602 IOL. The authors recommend careful selection of the IOL material for this technique to prevent possible complications.
Keywords: Haptic-Optic Junction, Intraocular Lens, Scleral Fixation, Yamane
References:
1. Yamane S, Sato S, Maruyama-Inoue M, Kadonosono K. Flanged intrascleral intraocular lens fixation with doubleneedle technique. Ophthalmology 2017;124:1136–1142.
2. Yalcinbayir O, Avci R, Ucan Gunduz G, Mavi Yildiz A, Cetin Efe A, Baykara M. Comparison of two techniques in posterior lens dislocations: Scleral suture fixation vs. modified Yamane intrascleral lens fixation. J Fr Ophtalmol 2022;45:13–19.
3. Fan KC, Patel NA, Al-Khersan H, Townsend JH. Outcomes of flanged intrascleral haptic fixation with CT Lucia threepiece intraocular lens. Ophthalmic Surg Lasers Imaging Retina 2021;52:658–662.
4. Muth DR, Wolf A, Kreutzer T, Shajari M, Vounotrypidis E, Priglinger S, et al. Safety and efficacy of current sclera fixation methods for intraocular lenses and literature overview. Klin Monbl Augenheilkd 2021;238:868–874.
5. Por YM, Chee SP. Spontaneous disinsertion of a multipiece foldable acrylic intraocular lens haptic 3 and 12 months after implantation. J Cataract Refract Surg 2004;30:1139–1142.
6. Fujishima K, Yoshitomi F, Oshika T. Spontaneous disinsertion of a haptic from a 3-piece acrylic foldable intraocular lens after surgery. J Cataract Refract Surg 2002;28:1296–1298.
7. Chuang CC, Lee JS, Lin KK, Hsiao CH. Late-onset spontaneous haptic disinsertion from a three-piece intraocular lens. Jpn J Ophthalmol 2010;54:500–502.
8. Solano JM, Baratz KH, Mahr MA, Erie JC. Late spontaneous haptic disinsertion from a three-piece intraocular lens. Am J Ophthalmol 2007;143:521–522.
9. Ma KK, Yuan A, Sharifi S, Pineda R. A Biomechanical study of flanged intrascleral haptic fixation of three-piece intraocular lenses. Am J Ophthalmol 2021;227:45–52.
10. Cheng CY, Chou YB, Tsai CY, Hsieh MH, Hsiao CC, Lai TT. Management of complications of sutureless intrascleral intraocular lens fixation. Taiwan J Ophthalmol 2024;14:95–101.
11. Stunf Pukl S, Kronschläger M, Ruiss M, Blouin S, Akcan ER, Findl O. Dislocation force of scleral flange-fixated intraocular lens haptics. BMC Ophthalmol 2024;24:103.
12. Stunf Pukl S, Kronschläger M, Ruiss M, Blouin S, Findl O. Size variability and fitting of 30-gauge thin-wall needles and 3-piece intraocular lens haptics for the flanged intrascleral fixation technique. J Cataract Refract Surg 2023;49:874–878.
13. Yannuzzi NA, Watane A, Patel NA, Townsend JH. Haptic and optic-haptic junction structural integrity of two 3-piece intraocular lenses. J Cataract Refract Surg 2022;48:743–744.
14. Rüfer F, Schröder A, Erb C. White-to-white corneal diameter: normal values in healthy humans obtained with the Orbscan II topography system. Cornea 2005;24:259–261.
15. Ganesh D, Lin SR. Global metrics on ocular biometry: Representative averages and standard deviations across ten countries from four continents. Eye (Lond) 2023;37:511–515.
16. Xu J, Lin P, Zhang S, Lu Y, Zheng T. Risk factors associated with intraocular lens decentration after cataract surgery. Am J Ophthalmol 2022;242:88–95.
17. Chen X, Gu X, Wang W, Xiao W, Jin G, Wang L, et al. Characteristics and factors associated with intraocular lens tilt and decentration after cataract surgery. J Cataract Refract Surg 2020;46:1126–1131.